BI-9000AT Digital Correlator for Photon Counting Applications
home enquiry profile products news contactr  
BI-200SM BI-MwA FOQELS BI-9000AT

Particle Size Analysis

Light Scattering

Centrifugal Sedimentation

Zeta Potential of Colloidal (Nano) Particles

Light Scattering

Accessories

Particle Size Standards and Microspheres

New! TurboCorr Digital Correlator

For photon counting applications including static and dynamic light scattering

Features at a glance

  • USB communications
  • Compact size, low power requirements, less than 300 mW
  • Auto- & cross-correlation modes
  • Up to 522 hardware channels
  • Sampling time 25 ns to 40 ms
  • Delay range 25 ns to 1,310 s
  • 100% efficient, real-time operation over the full delay time range
Supplied with control software for use with Windows™


The new TurboCorr digital correlator is the successor to the BI-9010AT and BI-9000AT. The new correlator card uses a cutting edge digital signal processor (DSP) which allows all the circuits to be embedded in a single, high density chip. The benefits of this design include full, in-system reprogrammability, significantly reduced power consumption as compared with earlier correlators and a compact design. The TurboCorr uses USB communications for enhanced portability and compatibility.

The TurboCorr is an advanced and extremely versatile digital photon correlator and counter designed for applications such as dynamic (DLS) and static (SLS) light scattering.

The TurboCorr has an adjustable delay range capable of producing a contiguous correlation function from 25 ns to 1,310 s representing a time axis that spans more than 10 decades. The middle and low speed circuits can be operated as independent correlators, simultaneously processing different input signals. This can be used, for example, to correlate two light scattering signals from different detector angles simultaneously. An on-board microprocessor on each correlator card controls the circuits and collects the correlation function(s). This relieves the CPU of the host computer of many tasks. Unlike other correlators, the TurboCorr never requires prescaling.

Operating modes

The TurboCorr digital correlator can be used for real-time operation in both auto- and cross-correlation modes. It is 100% efficient in either mode over the entire delay time range. Through the software, the user can select the number of hardware channels, either linearly spaced or geometrically spaced. In addition, channel spacing can be user-defined. The TurboCorr can be used for a wide variety of dynamic light scattering applications.

 

The TurboCorr is supplied complete with the BI-DLSW control software for Windows. In addition, Brookhaven offers a number of additional software programmes suitable for use with the TurboCorr and the BI-200SM laser light scattering goniometer system. These applications include particle sizing by dynamic light scattering (cumulants, NNLS, CONTIN, exponential sampling, and double exponential), characterization of gels and glass transitions (Williams-Watts), and molecular weight measurements by total intensity methods ( Zimm , Berry , and Debye plots). A user friendly alignment/stability test program (BI-IST) is available for use with the BI-200SM precision goniometer.



Channel layout :
Section Channels Sampling Time Range Maximum Time Delay Multiplier Bits
Low Speed 1 to 240 50 µs to 40 ms 1,310 s (40 ms) 16 x 16
Mid Speed 1 to 256 500 ns to 100 µs 1.6 s (100 µs) 8 x 8
Std High Speed 14 Channels at 100, 200, 300, 400, 500, 600, 800 ns, 1.0, 1.2, 1.4, 1.6, 2.0 and 2.4 µs
Opt High Speed 12 Channels at 25, 50, 75, 125, 150,175, 225, 250, 275, 325, 350 and 375 ns

BI-9000AT Specifications:
Format: USB 2.0. Four signal inputs and two low voltage outputs.
Maximum Number of Channels: 240 + 256 + 14 + 12 = 522
Delay Time Axis: User selectable (except fixed channels) from linear, constant ratio, or as individual time points recalled from user files.
Dynamic Range: Over 10 decades (25 ns to 1,310 seconds).
Operation Modes: Automatic and Manual. User definable macro commands to write, save, and recall unlimited number of sequences for single keystroke implementation.
Maximum Count Rate: 4 x 107 per second.
Standard Input Levels: TTL, 50 ohm input impedance, 5 ns minimum pulse width. Other levels via, optional, external converter.
Dimensions: 27 x 89 x 70 mm
Weight: 85 g
Power Requirements: 300 mW per card
Environmental Characteristics: Temperature 100C to 750C, Humidity 0% to 95%, non-condensing.

Options:
BI-25NS: 12 additional high speed (25 ns) channels.
BI-CON: Stepping motor controller, for use with the BI-200SM goniometer and accessories.
BI-ISTW: Alignment/Stability Software (for BI-200SM).
BI-ZPW: Zimm plot software for polymer molecular weights.
BI-ISDAW: Particle size distribution software suite.
BI-PSDW: Simple correlator control and particle size analysis programme.
BI-PSDW/NNLS: Adds NNLS algorithm to BI-PSDW.
BI-SLSW: Particle and fluid structure algorithms: Guinier, Kratky, Fractal, Rg, ln (Iex) vs q.

 

Updated 12.02.07

home | enquiry | profile | products |  contact